dc.description.abstractThe aim of this study was to evaluate the radiopacity of different gutta-percha points (Endo Points®, Dentsply®, Tanari®, Meta®, Roeko® and Odous®) in samples of 1 mm thick as established by ANSI/ADA Specification #57 and ISO 6876/2001, in comparison with thinner samples. Twelve test specimens for each material, four for each thickness (0.3, 0.6, and 1 mm and diameter of 8 mm), were laminated and compressed between two polished glass plates until the desirable thickness. Digital radiographs were obtained along with a graduated aluminum stepwedge varying from 1 to 10 mm in thickness. The X-ray unit was set at 70 kVp, 10 mA and 0.4 s exposure time, at a focal distance of 36 cm. One calibrated observer quantified the average values of pixels with Adobe Photoshop® software. Data were analyzed using ANOVA and Tukey tests, at 5% significance level. At 0.6 and 1 mm thickness, all the tested materials showed radiopacity higher than 3 mm of aluminum (reference value). At 0.3 mm thickness, Odous and Tanari presented significantly less radiopacity than the reference, and the other materials showed similar radiopacity to the reference. The study concluded that the materials demonstrated different radiopacities and all had values above the minimum recommended by ANSI/ ADA specification #57, being Odous and Tanari less radiopaque than the reference value in thinner samples (0.3mm).en
dc.format.mimetypeapplication/pdfpt_BR
dc.language.isoengpt_BR
dc.relation.ispartofBrazilian dental journal. Ribeirão Preto. Vol. 28, no. 5 (Sept./Oct. 2017), p. 592-596pt_BR
dc.rightsOpen Accessen
dc.subjectEndodontiapt_BR
dc.subjectEndodonticsen
dc.subjectGuttaperchaen
dc.subjectRadiologiapt_BR
dc.subjectInstrumentos odontológicospt_BR
dc.subjectRadiologyen
dc.subjectRadiopacityen
dc.titleRadiopacity evaluation of gutta-percha points in thinner samples than the ANSI/ADA recommendationpt_BR