Methods for vanadium determination in fuel oil by GF AAS with microemulsification and acid digestion sampling
dc.contributor.author | Amorim, Fábio A.C. | pt_BR |
dc.contributor.author | Lima, Daniel de Castro | pt_BR |
dc.contributor.author | Amaro, Joana Angélica de Azerêdo | pt_BR |
dc.contributor.author | Vale, Maria Goreti Rodrigues | pt_BR |
dc.contributor.author | Ferreira, Sérgio Luis Costa | pt_BR |
dc.date.accessioned | 2010-07-13T04:19:03Z | pt_BR |
dc.date.issued | 2007 | pt_BR |
dc.identifier.issn | 0103-5053 | pt_BR |
dc.identifier.uri | http://hdl.handle.net/10183/24582 | pt_BR |
dc.description.abstract | Dois métodos para determinação de vanádio em óleo combustível são propostos utilizando GF AAS. O primeiro é baseado na preparação de uma microemulsão da amostra que é diretamente injetada no tubo de grafite. As temperaturas de pirólise e atomização, 1500 e 2700 °C, respectivamente, foram otimizadas a partir das curvas de temperatura. Para este procedimento obteve-se um limite de detecção de 0,10 μg g-1 e uma massa característica de 87 pg. O outro método utiliza digestão da amostra com ácido sulfúrico, ácido nítrico e peróxido de hidrogênio. As temperaturas de pirólise e atomização, otimizadas, foram 1000 e 2700 °C, respectivamente. Para este procedimento, obteve-se um limite de detecção de 0,25 μg g-1 e uma massa característica de 63 pg. A exatidão de ambos os métodos foi confirmada pela análise do material de referência certificado NIST SRM 1634c-trace metals in fuel oil. | pt_BR |
dc.description.abstract | Two methods for vanadium determination in fuel oil are proposed using GF AAS. The first is based on the preparation of a microemulsion followed by direct injection into the graphite tube. The pyrolysis and atomization temperatures, 1500 and 2700 °C, respectively, were chosen from the temperature curves. This procedure allows determination of vanadium with a detection limit of 0.10 μg g-1 and characteristic mass of 87 pg. The other method uses total acid digestion of the sample with sulfuric acid, nitric acid and hydrogen peroxide. In this procedure, the pyrolysis and atomization temperatures recommended were 1000 and 2700 °C, respectively. Among the analytical characteristics for this method are a detection limit of 0.25 μg g-1 and a characteristic mass of 63 pg. The accuracy of both procedures was confirmed by the determination of vanadium in the certified reference material NIST SRM 1634c-trace metals in fuel oil. | en |
dc.format.mimetype | application/pdf | |
dc.language.iso | eng | pt_BR |
dc.relation.ispartof | Journal of the Brazilian Chemical Society. São Paulo. Vol. 18, n.8 (2007), p. 1566-1570 | pt_BR |
dc.rights | Open Access | en |
dc.subject | Vanadium | en |
dc.subject | Vanádio : Determinação | pt_BR |
dc.subject | Fuel oil | en |
dc.subject | Espectrometria de absorcao atomica em forno de grafite | pt_BR |
dc.subject | Microemulsoes | pt_BR |
dc.subject | Microemulsion | en |
dc.subject | Digestion sample | en |
dc.subject | GF AAS | en |
dc.title | Methods for vanadium determination in fuel oil by GF AAS with microemulsification and acid digestion sampling | pt_BR |
dc.type | Artigo de periódico | pt_BR |
dc.identifier.nrb | 000626188 | pt_BR |
dc.type.origin | Nacional | pt_BR |
Este item está licenciado na Creative Commons License
-
Artigos de Periódicos (40977)Ciências Exatas e da Terra (6198)