Avaliação da fadiga cíclica, torcional e deflexão angular de um novo instrumento endodôntico
dc.contributor.advisor | Rosa, Ricardo Abreu da | pt_BR |
dc.contributor.author | Silva, Giovana Siocheta da | pt_BR |
dc.date.accessioned | 2022-12-02T04:52:44Z | pt_BR |
dc.date.issued | 2021 | pt_BR |
dc.identifier.uri | http://hdl.handle.net/10183/252192 | pt_BR |
dc.description.abstract | O objetivo deste estudo foi avaliar a resistência à fadiga cíclica, torcional e deflexão angular de um novo instrumento endodôntico de níquel-titânio (NiTi) (Flat File 25.04) para preparo dos canais radiculares. Para os testes, foram selecionados 60 instrumentos: ProDesign Logic2 25.03 (n=20) e 25.05 (n=20) (Easy Equipamentos Odontológicos, Belo Horizonte, Brasil), e MK Flat File 25.04 (n=20) (MK Life, Porto Alegre, Brasil). Para o teste de fadiga cíclica, os instrumentos foram introduzidos em um canal artificial de aço inoxidável com uma angulação de 60º e um raio de curvatura de 5 mm localizado à 5 mm de sua ponta. O teste de fadiga cíclica foi realizado em temperatura ambiente controlada (37o ± 1 oC). O tempo e o número de ciclos para fratura foi gravado e calculado. O teste de torção foi baseado no protocolo ISO 3630-1, em que os 3 mm da ponta de cada instrumento foram fixados e conectados a um motor, um processador e uma célula de carga. O torque (N/cm) necessário para fraturar o instrumento foi obtido e a deflexão angular até o momento da fratura foi mensurada (em graus). Os dados foram analisados utilizando uma análise de variação unidirecional (ANOVA) e teste de Tukey com nível de significância de 5%. O instrumento Flat File 25.04 apresentou menor resistência à fadiga cíclica que os instrumentos Logic2 25.03 e 25.05 (P < 0,05). Flat File 25.04 apresentou torque similar ao Logic2 25.05 (P < 0,05) e ambos foram superiores ao Logic2 25.03 (P < 0,05). Os valores de deflexão angular foram diferentes para os três instrumentos testados (P < 0,05), em ordem decrescente conforme segue: Logic2 25.03, Logic2 25.05 e Flat File 25.04. As propriedades mecânicas dos instrumentos avaliados dependem das características geométricas dos instrumentos, tipo de liga de Ni-Ti utilizada, tratamento térmico da liga, processo de usinagem e fabricação dos instrumentos. O instrumento Flat File apresentou boas propriedades mecânicas após os testes de fadiga cíclica, torcional e deflexão angular. | pt_BR |
dc.description.abstract | The aim of this study was to evaluate the resistance to cyclic fatigue, torsional resistance and angular deflection of a new nickel-titanium (NiTi) (Flat File 25.04) instrument for root canal preparation. Sixty instruments were selected for the mechanical tests: ProDesign Logic2 25.03 (n=20) and 25.05 (n=20) (Easy Equipamentos Odontológicos, Belo Horizonte, Brazil), and MK Flat File 25.04 (n=20) (MK Life, Porto Alegre, Brazil). For the cyclic fatigue test, the instruments were introduced into an artificial stainless steel simulated canal with an angle of 60° and a radius of curvature of 5mm located 5mm from its tip. The cyclic fatigue test was performed at a controlled room temperature (37º ± 1ºC). The time and number of cycles to fracture were recorded and calculated. The torsional resistance test was based on the ISO 3630- 1 protocol, in which the 3mm tip of each instrument was fixed and connected to an electric motor, a processor and a load cell. The torque (N/cm) necessary to fracture the instrument was obtained and the angular deflection until the moment of fracture was measured (in degrees). Data were analyzed using a one-way analysis of variation (ANOVA) and Tukey's test with a significance level of 5%. Flat File 25.04 instrument had lower cyclic fatigue strength than the Logic 25.03 and 25.05 instruments (P < 0.05). Flat File 25.04 had similar torque to Logic2 25.05 (P < 0.05), and both were superior to Logic2 25.03 (P < 0.05). The angular deflection values were different for the three tested instruments (P < 0.05), as the following order: Logic2 25.03, Logic2 25.05 and Flat 25.04. The mechanical properties of the evaluated instruments depend on the geometric characteristics, Ni-Ti alloy, heat treatment of the alloy, machining and manufacturing process. Flat File instrument presented satisfactory mechanical properties after cyclic fatigue, torsional resistance and angular deflection tests. | en |
dc.format.mimetype | application/pdf | pt_BR |
dc.language.iso | por | pt_BR |
dc.rights | Open Access | en |
dc.subject | Endodontics | en |
dc.subject | Endodontia | pt_BR |
dc.subject | Cyclic fatigue | en |
dc.subject | Fadiga | pt_BR |
dc.subject | Torcional resistance | en |
dc.subject | NiTi instruments | en |
dc.title | Avaliação da fadiga cíclica, torcional e deflexão angular de um novo instrumento endodôntico | pt_BR |
dc.title.alternative | Evaluation of cyclic, torsional fatigue and angular deflection of a new endodontic instrument | en |
dc.type | Trabalho de conclusão de graduação | pt_BR |
dc.identifier.nrb | 001138940 | pt_BR |
dc.degree.grantor | Universidade Federal do Rio Grande do Sul | pt_BR |
dc.degree.department | Faculdade de Odontologia | pt_BR |
dc.degree.local | Porto Alegre, BR-RS | pt_BR |
dc.degree.date | 2021 | pt_BR |
dc.degree.graduation | Odontologia | pt_BR |
dc.degree.level | graduação | pt_BR |
Este item está licenciado na Creative Commons License
-
TCC Odontologia (982)